Analytical Method Validation of Selected Organochlorine Pesticides in Fortified Rice Using Gas Chromatography with Electron Capture Detector
Main Authors: | Wardani, Mutia Kusuma, Sutriyanto, Bambang, Lukitaningsih, Endang, Rohman, Abdul |
---|---|
Format: | Article info application/pdf eJournal |
Bahasa: | eng |
Terbitan: |
Fakultas Farmasi, Universitas Gadjah Mada
, 2013
|
Online Access: |
http://journal.ugm.ac.id/jfps/article/view/1859 http://journal.ugm.ac.id/jfps/article/view/1859/1666 |
Daftar Isi:
- Gas chromatography with electron capture detector (GC-ECD) has been validated for simultaneous determination of selected organochlorine pesticides namely heptachlor, endosulfan, dieldrin, dan p,p’-dichlorodiphenyltrichloroethane (p,p’-DDT) using aldrin as standard internal. Some parameters for analytical method validation were evaluated, i.e. linearity, precision, sensitivity expressed with limit of detection and limit of quantification, and recovery. GC-ECD was linear for determination of these pesticides with r values ≥ 0.99. The values of LOD obtained were 2.50, 2.97, 1.61, and 2.54 ng/mL for heptachlor, endosulphan, dieldrin, and p,p’-DDT, respectively. Furthermore, the relative standard deviation (RSD) values obtained for the assessment of analytical precision were lower than those required by Horwitz. Finally, the used GC-ECD method was accurate for determination of these organochlorine psticides in fortified rice samples which were indicated by the acceptable recovery percentages.