ctrlnum article-171
fullrecord <?xml version="1.0"?> <dc schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd"><title lang="en-US">TAC METHOD TO OVERCOME THE PRACTICAL DIFFICULTY IN THE CALIBRATION OF DWT WITH INSENSITIVE PISTON</title><creator>Ega, Adindra Vickar; Pusat Penelitian Metrologi - LIPI Research Center for Metrology - LIPI</creator><creator>Samodro, Rudi Anggoro; Pusat Penelitian Metrologi - LIPI Research Center for Metrology - LIPI</creator><subject lang="en-US">metrology; pressure metrology; measurement</subject><subject lang="en-US">CF method; TAC method; low accuracy DWT; sensitive; calibration .</subject><subject lang="en-US">pressure metrology</subject><description lang="en-US">The crossfloat (CF) method is well-known for its accurate and reliable results, however the difficulty on performing CF method arises when dealing with insensitive piston of calibrated DWT. In contrary, automatic transducer assisted crossfloat (TAC) method could be implemented since it does not affected by this condition as it advantages. The used of TAC method could avoid up to 0.008% relative error of effective area at zero pressure and at 20 &#xB0;C (A0,20) and twice higher error distortion coefficient (b) in case of calibrating insensitive piston of DWT. In addition, the TAC method has another advantage that is save time with consistent result because it can be performed automatically. Once should be consider when using TAC method is the accuracy of pressure transducer. In this case study, the existing pressure transducer owned by RCM &#x2013; LIPI, could be utilized to perform automatic TAC method in lower pressure range down to 2 MPa that below 10% of its maximum scale, to calibrate low accuracy DWT greater than 0.08%.</description><publisher lang="en-US">LIPI Press, Anggota IKAPI</publisher><contributor lang="en-US"/><date>2019-11-11</date><type>Journal:Article</type><type>Other:info:eu-repo/semantics/publishedVersion</type><type>Other:</type><type>Other:TAC method</type><type>File:application/pdf</type><identifier>http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/view/171</identifier><identifier>10.31153/instrumentasi.v43i1.171</identifier><source lang="en-US">Instrumentasi; Vol 43, No 1 (2019); 11-23</source><source lang="id-ID">Instrumentasi; Vol 43, No 1 (2019); 11-23</source><source>2460-1462</source><source>0125-9202</source><language>eng</language><relation>http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/view/171/pdf_48</relation><relation>http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/downloadSuppFile/171/50</relation><relation>http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/downloadSuppFile/171/51</relation><rights lang="en-US">Copyright (c) 2019 Instrumentasi</rights><rights lang="en-US">http://creativecommons.org/licenses/by-nc/4.0</rights><recordID>article-171</recordID></dc>
language eng
format Journal:Article
Journal
Other:info:eu-repo/semantics/publishedVersion
Other
Other:
Other:TAC method
File:application/pdf
File
Journal:Journal
author Ega, Adindra Vickar; Pusat Penelitian Metrologi - LIPI Research Center for Metrology - LIPI
Samodro, Rudi Anggoro; Pusat Penelitian Metrologi - LIPI Research Center for Metrology - LIPI
title TAC METHOD TO OVERCOME THE PRACTICAL DIFFICULTY IN THE CALIBRATION OF DWT WITH INSENSITIVE PISTON
publisher LIPI Press, Anggota IKAPI
publishDate 2019
topic metrology
pressure metrology
measurement
CF method
TAC method
low accuracy DWT
sensitive
calibration
url http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/view/171
http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/view/171/pdf_48
http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/downloadSuppFile/171/50
http://jurnalinstrumentasi.bsn.go.id/index.php/ji/article/downloadSuppFile/171/51
contents The crossfloat (CF) method is well-known for its accurate and reliable results, however the difficulty on performing CF method arises when dealing with insensitive piston of calibrated DWT. In contrary, automatic transducer assisted crossfloat (TAC) method could be implemented since it does not affected by this condition as it advantages. The used of TAC method could avoid up to 0.008% relative error of effective area at zero pressure and at 20 °C (A0,20) and twice higher error distortion coefficient (b) in case of calibrating insensitive piston of DWT. In addition, the TAC method has another advantage that is save time with consistent result because it can be performed automatically. Once should be consider when using TAC method is the accuracy of pressure transducer. In this case study, the existing pressure transducer owned by RCM – LIPI, could be utilized to perform automatic TAC method in lower pressure range down to 2 MPa that below 10% of its maximum scale, to calibrate low accuracy DWT greater than 0.08%.
id IOS15518.article-171
institution Badan Standardisasi Nasional
institution_id 258
institution_type library:special
library
library Perpustakaan Badan Standardisasi Nasional
library_id 22
collection Jurnal Instrumentasi
repository_id 15518
subject_area Standardization/Standarisasi
Instrumentation
Conformity Assessment
city JAKARTA PUSAT
province DKI JAKARTA
repoId IOS15518
first_indexed 2021-07-13T02:11:13Z
last_indexed 2021-07-13T02:11:13Z
recordtype dc
_version_ 1705133886631051264
score 17.610468