MODIFIKASI NEBULIZER KOMPRESOR DENGAN MENAMBAHKAN PENGATURAN TIMER DAN DETEKTOR CAIRAN OBAT SEBAGAI BATASAN WAKTU TERAPI PEMBERIAN OBAT PADA PENDERITA ASMA

Main Author: Fernando, Alex Surapati, Faisal Hadi, Andica
Format: Article info application/pdf eJournal
Bahasa: eng
Terbitan: University of Bengkulu , 2016
Subjects:
Online Access: https://ejournal.unib.ac.id/index.php/teknosia/article/view/1021
https://ejournal.unib.ac.id/index.php/teknosia/article/view/1021/848
ctrlnum article-1021
fullrecord <?xml version="1.0"?> <dc schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd"><title lang="en-US">MODIFIKASI NEBULIZER KOMPRESOR DENGAN MENAMBAHKAN PENGATURAN TIMER DAN DETEKTOR CAIRAN OBAT SEBAGAI BATASAN WAKTU TERAPI PEMBERIAN OBAT PADA PENDERITA ASMA</title><creator>Fernando, Alex Surapati, Faisal Hadi, Andica</creator><subject lang="en-US">asthma; modification; nebulizer compressor; therapy</subject><description lang="en-US">Modifications of&#xA0; nebulizer compressor in this study were made and can be used to solve problems in the treatment of asthma therapy. Asthma is a symptom that is caused by abnormalities of the respiratory tract, causing increased sensitivity to environmental stimuli. Symptoms are often seen in people with asthma usually is cold / sneezing, coughing, itchy throat. Nebulizer is a device used in the treatment of asthma. This tool can transform the drug particles from a liquid into a very fine grains (steam). Because the drug is converted into a gas (steam), the drug is easier to be absorbed so that the effects of the drug seem faster than oral medications. Overall the nebulizer has worked and the functioning properly fit the desired time and the limits of medicine. The output voltage sensor of potions detector around 0.726 to 1.221 volts with a range of drug volumes 1-6 cc. Overall the percentage of error in testing time is between 6:49%. The average time required for evaporation of the drug in 1 cc of liquid drug is approximately 638.86 seconds.</description><publisher lang="en-US">University of Bengkulu</publisher><contributor lang="en-US"/><date>2016-09-01</date><type>Journal:Article</type><type>Other:info:eu-repo/semantics/publishedVersion</type><type>Other:</type><type>File:application/pdf</type><identifier>https://ejournal.unib.ac.id/index.php/teknosia/article/view/1021</identifier><source lang="en-US">Teknosia; Vol 2, No 17 (2016): X, September; 1-11</source><source>1978-8819</source><language>eng</language><relation>https://ejournal.unib.ac.id/index.php/teknosia/article/view/1021/848</relation><recordID>article-1021</recordID></dc>
language eng
format Journal:Article
Journal
Other:info:eu-repo/semantics/publishedVersion
Other
Other:
File:application/pdf
File
Journal:eJournal
author Fernando, Alex Surapati, Faisal Hadi, Andica
title MODIFIKASI NEBULIZER KOMPRESOR DENGAN MENAMBAHKAN PENGATURAN TIMER DAN DETEKTOR CAIRAN OBAT SEBAGAI BATASAN WAKTU TERAPI PEMBERIAN OBAT PADA PENDERITA ASMA
publisher University of Bengkulu
publishDate 2016
topic asthma
modification
nebulizer compressor
therapy
url https://ejournal.unib.ac.id/index.php/teknosia/article/view/1021
https://ejournal.unib.ac.id/index.php/teknosia/article/view/1021/848
contents Modifications of nebulizer compressor in this study were made and can be used to solve problems in the treatment of asthma therapy. Asthma is a symptom that is caused by abnormalities of the respiratory tract, causing increased sensitivity to environmental stimuli. Symptoms are often seen in people with asthma usually is cold / sneezing, coughing, itchy throat. Nebulizer is a device used in the treatment of asthma. This tool can transform the drug particles from a liquid into a very fine grains (steam). Because the drug is converted into a gas (steam), the drug is easier to be absorbed so that the effects of the drug seem faster than oral medications. Overall the nebulizer has worked and the functioning properly fit the desired time and the limits of medicine. The output voltage sensor of potions detector around 0.726 to 1.221 volts with a range of drug volumes 1-6 cc. Overall the percentage of error in testing time is between 6:49%. The average time required for evaporation of the drug in 1 cc of liquid drug is approximately 638.86 seconds.
id IOS2605.article-1021
institution Universitas Bengkulu
institution_id 179
institution_type library:university
library
library Perpustakaan Universitas Bengkulu
library_id 477
collection Teknosia
repository_id 2605
city KOTA BENGKULU
province BENGKULU
shared_to_ipusnas_str 1
repoId IOS2605
first_indexed 2017-02-25T14:51:06Z
last_indexed 2018-09-17T09:26:28Z
recordtype dc
_version_ 1685801017445187584
score 17.611225