fullrecord <?xml version="1.0"?> <dc schemaLocation="http://www.openarchives.org/OAI/2.0/oai_dc/ http://www.openarchives.org/OAI/2.0/oai_dc.xsd"><title lang="en-US">The Preliminary Prototype of Medium Dose Rate Brachytherapy Equipment</title><creator>Satmoko, A.; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313</creator><creator>Harjanto, T.; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313</creator><creator>Putra, I.M.; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313</creator><creator>Kristiyanti, Kristiyanti; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313</creator><subject lang="en-US">Brachytherapy; Prototype; Design; Construction; Medium dose rate</subject><description lang="en-US">A preliminary prototype of a brachytherapy equipment has been constructed. The work started by developing conceptual design, followed by basic design and detailed design. In the conceptual design, design requirements are stated. In the basic design, technical specifications for main components are determined. In detailed design, general drawings are discussed. The prototype consists of three main systems: a mechanical system, an instrumentation system, and a safety system. The mechanical system assures the movement mechanism of the isotope source position beginning from the standby position until the applicators. It consists of three main modules: a position handling module, a container module, and a channel distribution module. The position handling module serves to move the isotope source position. As shielding, the second module is to store the source when the equipment is in standby position. The prototype provides 12 output channels. &#xA0;The channel selection is performed by the third module. The instrumentation system controls the movement of source position by handling motor operations. It consists of several modules. A microcontroller module serves as a control center whose task includes both controlling motors and communicating with computer. A motor module serves to handle motors. 10 sensors, including their signal conditionings, are introduced to read the environment conditions of the equipment. LEDs are used to display these conditions. In order to facilitate the operators&#x2019; duty, communication via RS232 is provided. The brachytherapy equipment can therefore be operated by using computer. Interface software is developed using C# language. To complete both mechanical and instrumentation systems performance, a safety system is developed to make sure that the safety for operator and patients from receiving excessive radiation. An interlock system is introduced to guard against abnormal conditions. In the worst case, a manual intervention by the operator is provided when all other means are failing to store the isotope source into the safe container. The tests showed good results. The prototype can send the isotope source to applicators. The isotope source can be positioned with an accuracy of&#xA0; &#xB1; 0.5 mm and with a speed of 550 mm/second. These characteristics meet the design criteria.Received: 16 January 2013; Revised: 26 August 2013; Accepted: 29 August 2013</description><publisher lang="en-US">PPIKSN-BATAN</publisher><contributor lang="en-US">National Nuclear Energy Agency, Indonesia</contributor><date>2013-10-03</date><type>Journal:Article</type><type>Other:info:eu-repo/semantics/publishedVersion</type><type>Other:</type><type>File:application/pdf</type><identifier>http://aij.batan.go.id/index.php/aij/article/view/234</identifier><identifier>10.17146/aij.2013.234</identifier><source lang="en-US">Atom Indonesia; Vol 39, No 2 (2013): August 2013; 75-81</source><source>2356-5322</source><source>0126-1568</source><language>eng</language><relation>http://aij.batan.go.id/index.php/aij/article/view/234/191</relation><rights lang="0">Copyright (c) 2016 Atom Indonesia</rights><rights lang="0">http://creativecommons.org/licenses/by-nc-sa/4.0</rights><recordID>article-234</recordID></dc>
language eng
format Journal:Article
Journal
Other:info:eu-repo/semantics/publishedVersion
Other
Other:
File:application/pdf
File
Journal:eJournal
author Satmoko, A.; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313
Harjanto, T.; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313
Putra, I.M.; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313
Kristiyanti, Kristiyanti; Center for Nuclear Instrumentation Engineering, National Nuclear Energy Agency, Puspiptek Area, Serpong, Tangerang 15313
author2 National Nuclear Energy Agency, Indonesia
title The Preliminary Prototype of Medium Dose Rate Brachytherapy Equipment
publisher PPIKSN-BATAN
publishDate 2013
topic Brachytherapy
Prototype
Design
Construction
Medium dose rate
url http://aij.batan.go.id/index.php/aij/article/view/234
http://aij.batan.go.id/index.php/aij/article/view/234/191
contents A preliminary prototype of a brachytherapy equipment has been constructed. The work started by developing conceptual design, followed by basic design and detailed design. In the conceptual design, design requirements are stated. In the basic design, technical specifications for main components are determined. In detailed design, general drawings are discussed. The prototype consists of three main systems: a mechanical system, an instrumentation system, and a safety system. The mechanical system assures the movement mechanism of the isotope source position beginning from the standby position until the applicators. It consists of three main modules: a position handling module, a container module, and a channel distribution module. The position handling module serves to move the isotope source position. As shielding, the second module is to store the source when the equipment is in standby position. The prototype provides 12 output channels. The channel selection is performed by the third module. The instrumentation system controls the movement of source position by handling motor operations. It consists of several modules. A microcontroller module serves as a control center whose task includes both controlling motors and communicating with computer. A motor module serves to handle motors. 10 sensors, including their signal conditionings, are introduced to read the environment conditions of the equipment. LEDs are used to display these conditions. In order to facilitate the operators’ duty, communication via RS232 is provided. The brachytherapy equipment can therefore be operated by using computer. Interface software is developed using C# language. To complete both mechanical and instrumentation systems performance, a safety system is developed to make sure that the safety for operator and patients from receiving excessive radiation. An interlock system is introduced to guard against abnormal conditions. In the worst case, a manual intervention by the operator is provided when all other means are failing to store the isotope source into the safe container. The tests showed good results. The prototype can send the isotope source to applicators. The isotope source can be positioned with an accuracy of ± 0.5 mm and with a speed of 550 mm/second. These characteristics meet the design criteria.Received: 16 January 2013; Revised: 26 August 2013; Accepted: 29 August 2013
id IOS96.article-234
institution Badan Tenaga Nuklir Nasional
institution_id 54
institution_type library:special
library
library Perpustakaan Badan Tenaga Nuklir Nasional
library_id 23
collection Atom Indonesia Journal
repository_id 96
subject_area Rekayasa
city JAKARTA SELATAN
province DKI JAKARTA
repoId IOS96
first_indexed 2016-09-24T07:11:47Z
last_indexed 2016-09-24T07:11:47Z
recordtype dc
_version_ 1722431331199090688
score 17.610487